Options
M02.006 Quantitative photometric determination of iron after derivatisation
Citation Link: https://doi.org/10.15480/336.2194
Type
Text
Date Issued
2017-11-27
Language
English
Institute
TORE-URI
Abstract
Determination of Fe (II) is carried out by means of 1,10-phenanthroline taking advantage of the fact that Fe (II) forms a water soluble red complex with 1,10-phenanthroline. The intensity of the coloured species is measured by a photometer. The concentration in samples is quantified with an external calibration and results are reported in mg/l.
Subjects
Analytical method
iron
DDC Class
540: Chemie
Loading...
Name
M02.006 - Determination of Iron.pdf
Size
235.19 KB
Format