Osteolysis following PE wear of a Hastings head on a monoblock hip stem
Citation Link: https://doi.org/10.15480/882.4119
Publikationstyp
Journal Article
Date Issued
2021-10-07
Sprache
English
Author(s)
Herausgeber*innen
Institut
TORE-DOI
Journal
Volume
2021
Start Page
1
End Page
7
Article Number
9989395
Citation
Case Reports in Orthopedics : 9989395 (2021)
Publisher DOI
Publisher
Hindawi
We report a case of extended osteolysis, requiring a third revision of the left hip in an 85-year-old man 46 years after index operation. Major polyethylene (PE) wear occurred due to a missmatched combination of a bipolar Hastings head with a PE liner and head damage of the originally maintained stem. This case demonstrates that bipolar heads should not be used with PE cup liners since the respective bearing diameters cannot be guaranteed to match due to missing specifications. Furthermore, putting a Hastings head on an already damaged head of the stem should be omitted and rather the stem should initially be revised.
DDC Class
570: Biowissenschaften, Biologie
Funding(s)
More Funding Information
We acknowledge the support for the open access fees from the Hamburg University of Technology (TUHH) in the
funding programme Open Access Publishing.