Absence of surface roughness changes after insertion of one type of matte cemented femoral component during 2 to 15 years

Publikationstyp
Journal Article
Date Issued
2003
Sprache
English
Author(s)
Knoch, Marius von  
Bluhm, Andrea  
Biomechanik M-3  
Morlock, Michael  
Biomechanik M-3  
Foerstér, Götz von
TORE-URI
https://hdl.handle.net/11420/46686
Journal
Journal of arthroplasty  
Volume
18
Issue
4
Start Page
471
End Page
477
Citation
Journal of Arthroplasty 18 (4): 471-477 (2003)
Publisher DOI
10.1016/S0883-5403(03)00062-7
Scopus ID
2-s2.0-0038266838
Publisher
Churchill Livingstone
Theoretically, micromotion might occur even in stable cement-implant interface conditions. It was hypothesized that smoothing of the matte surfaces of femoral components because of abrasion or corrosion phenomena is a possible consequence of micromotion. Two new and 11 femoral components that were retrieved after 2 to 15 years of stable cement-implant interface conditions were inspected, investigated with light microscopy, and evaluated with surface roughness measurement by a profilometer. No surface changes were observed such as abrasion or corrosion phenomena except at the tip and the collar of the femoral components. The surface roughness was Ra = 1.0 and 1.1 μm in the new femoral components and Ra (mean) = 0.96 μm (range: 0.8-1.5 μm) in the retrieved components. There was no correlation between time in situ and surface roughness as confirmed by regression analysis. These data suggest that significant micromotion was not present at the studied interface.
Subjects
Arthroplasty
Cemented
Hip
Roughness
Surface
DDC Class
610: Medicine, Health
TUHH
Weiterführende Links
  • Contact
  • Send Feedback
  • Cookie settings
  • Privacy policy
  • Impress
DSpace Software

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science
Design by effective webwork GmbH

  • Deutsche NationalbibliothekDeutsche Nationalbibliothek
  • ORCiD Member OrganizationORCiD Member Organization
  • DataCiteDataCite
  • Re3DataRe3Data
  • OpenDOAROpenDOAR
  • OpenAireOpenAire
  • BASE Bielefeld Academic Search EngineBASE Bielefeld Academic Search Engine
Feedback