Please use this identifier to cite or link to this item: https://doi.org/10.15480/882.1510
Fulltext available Open Access
Publisher DOI: 10.2316/P.2016.832-042
Title: Wireless blood pressure measurement implant electronics for integration in a stent craft
Language: English
Authors: John, Bibin 
Ranjan, Rajeev 
Spink, Clemens 
Schröder, Dietmar 
Koops, Andreas 
Adam, Gerhard 
Krautschneider, Wolfgang 
Keywords: telemetry;wireless power transmission;inductive coupling;Implantat;pressure measurement
Issue Date: Feb-2016
Abstract (english): This paper describes the design, implementation and testing of a wireless blood pressure sensing implant. The implant is mounted on a stent graft. The implant along with the stent graft must fit into a 16F applicator stent delivery system. The implant is powered using the energy harvested from wireless power transmission. The measured pressure data is sent wirelessly to an external data reader hardware. The pressure measurement must have resolution of 1 mmHg.
Conference: The 12th IASTED International Conference on Biomedical Engineering BioMed 2016 
URI: http://tubdok.tub.tuhh.de/handle/11420/1513
DOI: 10.15480/882.1510
ISBN: 978-0-88986-981-3
978-0-88986-980-6
Institute: Nano- und Medizinelektronik E-9 
Type: InProceedings (Aufsatz / Paper einer Konferenz etc.)
License: In Copyright In Copyright
Appears in Collections:Publications with fulltext