Gladiß, Anselm vonAnselm vonGladißGräser, MatthiasMatthiasGräserBuzug, Thorsten M.Thorsten M.Buzug2019-11-282019-11-282018-03Workshop on Bildverarbeitung fur die Medizin, 2018: 92-97978-3-540-29595-2978-3-540-29594-5http://hdl.handle.net/11420/3907In Magnetic Particle Imaging, superparamagnetic iron oxide nanoparticles are excited by an oscillating magnetic field that is generated by sending coils. In multidimensional set-ups, perfect geometrical decoupling cannot be achieved. Remaining coupling can distort the trajectory and introduce artefacts, thus active decoupling is introduced at costs of high power consumption. This work uses a device with active coupling control to investigate the influence of different coupling levels on the image reconstruction.enTechnology::600: TechnologyInfluence of excitation signal coupling on reconstructed images in magnetic particle imagingConference Paper10.1007/978-3-662-56537-7_36Conference Paper