Options
A Slim Needle Neural Probe with 160 Active Recording Sites and Selectable ADCs
Publikationstyp
Conference Paper
Publikationsdatum
2019-10
Sprache
English
Institut
TORE-URI
Article Number
8919180
Citation
Biomedical Circuits and Systems Conference (BioCAS): 8919180 (2019-10)
Contribution to Conference
Publisher DOI
Scopus ID
In order to provide active neural needle probes with a large number of recordings sites integrated on a single shank, a high degree of miniaturization of the probe shank and its electronics is required. This work presents a probe which comprises 160 recording sites each combined with a low-noise amplifier integrated under each electrode in an area of only 50×80 μm2. In total, 32 channels can be monitored simultaneously by means of 32 12-bit incremental Delta-Sigma ADCs each requiring an area of 120×40 μm2. The selection of the 32 channels is performed by means of a digital controller. The average input referred noise for AP signals (300 Hz to 10 kHz) is extracted to 20.87 μVrms in in-vitro measurements. The design competes well with the state of the art due to a maximum shank width of 200 μm, an area of 0.032 mm2, and a power consumption of 26.5 μW per channel.
Schlagworte
active neural recording
Area-efficient electronics
incremental Delta-Sigma ADCs
slim neural probe