  4. Interfacial conditions between a press-fit acetabular cup and bone during daily activities : implications for achieving bone in-growth
 
Options

Interfacial conditions between a press-fit acetabular cup and bone during daily activities : implications for achieving bone in-growth

Publikationstyp
Journal Article
Date Issued
2000-11-01
Sprache
English
Author(s)
Spears, Iain R.  
Pfleiderer, Martin
Schneider, Erich  
Hille, Ekkehard  
Bergmann, Georg
Morlock, Michael  
Biomechanik M-3  
TORE-URI
https://hdl.handle.net/11420/46619
Journal
Journal of biomechanics  
Volume
33
Issue
11
Start Page
1471
End Page
1477
Citation
Journal of Biomechanics 33 (11): 1471-1477 (2000)
Publisher DOI
10.1016/S0021-9290(00)00096-8
Scopus ID
2-s2.0-0034333770
Publisher
Elsevier Science
Interfacial gaps and relative micromotions during activities are widely believed to restrict the boney in-growth process of non-cemented acetabular cups. Using finite element modeling of the cup-bone system, relative micromotions and interfacial gaps are calculated for walking slow, normal and fast and for climbing upstairs, downstairs and standing up from a chair. A 2mm press-fit is simulated and interfacial conditions in the immediate postoperative period (i.e. prior to boney in-growth) calculated between paired nodes covering the whole of the interface. In regions of 'safe' micromotions and 'allowable' gaps, boney in-growth is simulated by specifying zero relative displacement between nodal pairs. The modified model is then resubjected to the loads associated with climbing upstairs, which was shown to be the worst activity. Interfacial conditions are recalculated for subsequent iterations. The procedure is repeated until no further in-growth is predicted. The final pattern of in-growth calculated with the model compares reasonably well with histological evidence from explanted canine cups (Cha et al., 1998. Transactions of the Orthopaedic Research Society, 23, p. 373). Bridging between adjacent regions of in-growth is observed. Notably, in-growth occurs at most of the periphery but not in the polar region. The lack of polar in-growth is caused by the interfacial gap assumed to exist after cup implantation. It is suggested that increasing/decreasing hip-joint loads would have little effect on this lack of polar in-growth. However, excessive micromotions as a result of high hip-joint loads cause a lack of in-growth in the anterior region of the periphery in the model. Although such results were not found in the canine study, if relevant to the general human population, the avoidance of harsh weight-bearing activities may encourage complete peripheral in-growth but is speculated to do little for polar in- growth.
Subjects
Acetabular cups
Bone in-growth
Finite elements
Primary stability
DDC Class
610: Medicine, Health
TUHH
Weiterführende Links
  • Contact
  • Send Feedback
  • Cookie settings
  • Privacy policy
  • Impress
DSpace Software

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science
Design by effective webwork GmbH

  • Deutsche NationalbibliothekDeutsche Nationalbibliothek
  • ORCiD Member OrganizationORCiD Member Organization
  • DataCiteDataCite
  • Re3DataRe3Data
  • OpenDOAROpenDOAR
  • OpenAireOpenAire
  • BASE Bielefeld Academic Search EngineBASE Bielefeld Academic Search Engine
Feedback