A versatile miniature bioreactor and its application to bioelectrochemistry studies

Publikationstyp
Journal Article
Date Issued
2010-04-21
Sprache
English
Author(s)
Kloke, Arne  
Rubenwolf, Stefanie  
Buck, Britta  
Gescher, Johannes  
Kerzenmacher, Sven  
Zengerle, Roland  
von Stetten, Felix  
TORE-URI
http://hdl.handle.net/11420/14785
Journal
Biosensors and bioelectronics  
Volume
25
Issue
12
Start Page
2559
End Page
2565
Citation
Biosensors and Bioelectronics 25 (12): 2559-2565 (2010-08-01)
Publisher DOI
10.1016/j.bios.2010.04.014
Scopus ID
2-s2.0-77951119622
PubMed ID
20537883
Publisher
Elsevier Science
Often, reproducible investigations on bio-microsystems essentially require a flexible but well-defined experimental setup, which in its features corresponds to a bioreactor. We therefore developed a miniature bioreactor with a volume in the range of a few millilitre that is assembled by alternate stacking of individual polycarbonate elements and silicone gaskets. All the necessary supply pipes are incorporated as bore holes or cavities within the individual elements. Their combination allows for a bioreactor assembly that is easily adaptable in size and functionality to experimental demands. It allows for controlling oxygen transfer as well as the monitoring of dissolved oxygen concentration and pH-value. The system provides access for media exchange or sterile sampling. A mass transfer coefficient for oxygen (kLa) of 4.3×10-3s-1 at a flow rate of only 15mlmin-1 and a mixing time of 1.5s at a flow rate of 11mlmin-1 were observed for the modular bioreactor. Single reactor chambers can be interconnected via ion-conductive membranes to form a two-chamber test setup for investigations on electrochemical systems such as fuel cells or sensors. The versatile applicability of this modular and flexible bioreactor was demonstrated by recording a growth curve of Escherichia coli (including monitoring of pH and oxygen) saturation, and also as by two bioelectrochemical experiments. In the first electrochemical experiment the use of the bioreactor enabled a direct comparison of electrode materials for a laccase-catalyzed oxygen reduction electrode. In a second experiment, the bioreactor was utilized to characterize the influence of outer membrane cytochromes on the performance of Shewanella oneidensis in a microbial fuel cell.
Subjects
Bioelectrochemical testing
Biofuel cell electrode characterization
Experimental setup for bioelectrochemistry
Modular miniature bioreactor
DDC Class
570: Biowissenschaften, Biologie
TUHH
Weiterführende Links
  • Contact
  • Send Feedback
  • Cookie settings
  • Privacy policy
  • Impress
DSpace Software

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science
Design by effective webwork GmbH

  • Deutsche NationalbibliothekDeutsche Nationalbibliothek
  • ORCiD Member OrganizationORCiD Member Organization
  • DataCiteDataCite
  • Re3DataRe3Data
  • OpenDOAROpenDOAR
  • OpenAireOpenAire
  • BASE Bielefeld Academic Search EngineBASE Bielefeld Academic Search Engine
Feedback