Publikationstyp
Journal Article
Publikationsdatum
2014-03
Sprache
English
Enthalten in
Volume
36
Issue
3
Start Page
300
End Page
307
Citation
Medical Engineering and Physics 36 (3): 300-307 (2014)
Publisher DOI
Scopus ID
Publisher
Elsevier
Several bi-modular hip prostheses exhibit an elevated number of fretting-related postoperative complications most probably caused by excessive micromotions at taper connections. This study investigated micromotions at the stem-neck interface of two different designs: one design (Metha, Aesculap AG) has demonstrated a substantial number of in vivo neck fractures for Ti-Ti couplings, but there are no documented fractures for Ti-CoCr couplings. Conversely, for a comparable design (H-Max M, Limacorporate) with a Ti-Ti coupling only one clinical failure has been reported. Prostheses were mechanically tested and the micromotions were recorded using a contactless measurement system.For Ti-Ti couplings, the Metha prosthesis showed a trend towards higher micromotions compared to the H-Max M (6.5. ±. 1.6. μm vs. 3.6. ±. 1.5. μm, p=0.08). Independent of the design, prostheses with Ti neck adapter caused significantly higher interface micromotions than those with CoCr ones (5.1. ±. 2.1. μm vs. 0.8. ±. 1.6. μm, p=0.001). No differences in micromotions between the Metha prosthesis with CoCr neck and the H-Max M with Ti neck were observed (2.6. ±. 2.0. μm, p=0.25).The material coupling and the design are both crucial for the micromotions magnitude. The extent of micromotions seems to correspond to the number of clinically observed fractures and confirm the relationship between those and the occurrence of fretting corrosion. © 2013 IPEM.
Schlagworte
Bi-modular prosthesis
Fretting
Micromotion
Neck fracture
Stem-neck interface
DDC Class
610: Medicine, Health
620: Engineering