Options
A 130NM ASIC for EMG signal acquisition to control a hand prosthetic
Publikationstyp
Conference Paper
Publikationsdatum
2012
Sprache
English
Start Page
149
End Page
153
Citation
BIODEVICES 2012 - Proceedings of the International Conference on Biomedical Electronics and Devices
Contribution to Conference
Scopus ID
Publisher
SciTePress
ISBN
978-989-8425-91-1
A mixed-signal chip in 130nm Technology is described. The chip acquires EMG signals from five differential inputs. The signals are amplified, multiplexed and digitized inside the developed chip. The ASIC is optimised for low power and low noise and is intended for application in medical implants. Furthermore, the chip provides a configuration options for several parameters for the analogue part of the chip. The chip has been tested by recording EMG signals from implanted Myo-Electrodes.
Schlagworte
Bio-signal amplifier
EMG signal acquisition
Hand prosthetic
Medical implant
Mixed-signal chip
DDC Class
610: Medicine, Health
621.3: Electrical Engineering, Electronic Engineering