Orientation of the acetabular component : a comparison of five navigation systems with conventional surgical technique

Publikationstyp
Journal Article
Date Issued
2006-10-01
Sprache
English
Author(s)
Honl, Matthias  
Biomechanik M-3  
Schwieger, Karsten  
Salineros, Matías J.
Jacobs, Joshua J.  
Morlock, Michael  
Biomechanik M-3  
Wimmer, Markus A.  
TORE-URI
https://hdl.handle.net/11420/46630
Journal
Journal of bone and joint surgery - BRITISH VOLUME  
Volume
88
Issue
10
Start Page
1401
End Page
1405
Citation
Journal of Bone and Joint Surgery, Series B 88 (10): 1401-1406 (2006)
Publisher DOI
10.1302/0301-620X.88B10.17587
Scopus ID
2-s2.0-33750367795
Publisher
Ed. Soc. of Bone and Joint Surgery
We compared the orientation of the acetabular component obtained by a conventional manual technique with that using five different navigation systems. Three surgeons carried out five implantations of an acetabular component with each navigation system, as well as manually, using an anatomical model. The orientation of the acetabular component, including inclination and anteversion, and its position was determined using a co-ordinate measuring machine. The variation of the orientation of the acetabular component was higher in the conventional group compared with the navigated group. One experienced surgeon took significantly less time for the procedure. However, his placement of the component was no better than that of the less experienced surgeons. Significantly better inclination and anteversion (p < 0.001 for both) were obtained using navigation. These parameters were not significantly different between the surgeons when using the conventional technique (p = 0.966). The use of computer navigation helps a surgeon to orientate the acetabular component with less variation regarding inclination and anteversion. © 2006 British Editorial Society of Bone and Joint Surgery.
DDC Class
610: Medicine, Health
620: Engineering
TUHH
Weiterführende Links
  • Contact
  • Send Feedback
  • Cookie settings
  • Privacy policy
  • Impress
DSpace Software

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science
Design by effective webwork GmbH

  • Deutsche NationalbibliothekDeutsche Nationalbibliothek
  • ORCiD Member OrganizationORCiD Member Organization
  • DataCiteDataCite
  • Re3DataRe3Data
  • OpenDOAROpenDOAR
  • OpenAireOpenAire
  • BASE Bielefeld Academic Search EngineBASE Bielefeld Academic Search Engine
Feedback