Options
Influence of the system matrix on channel leakage artifacts in multi-contrast MPI
Citation Link: https://doi.org/10.15480/882.9520
Publikationstyp
Journal Article
Date Issued
2024-03-10
Sprache
English
TORE-DOI
Volume
10
Issue
1, suppl. 1
Article Number
2403033
Citation
International Journal on Magnetic Particle Imaging 10 (1, suppl. 1): 2403033 (2024)
Publisher DOI
Scopus ID
Publisher
Infinite Science Publishing
Magnetic Particle Imaging (MPI) is a tracer-based medical imaging modality with great potential due to its high sensitivity, high spatiotemporal resolution, and ability to quantify the tracer concentration. Image reconstruction in MPI is an ill-posed problem, which can be addressed by the use of regularization methods. Single-and multi-contrast MPI reconstructions produce different kinds of artifacts. In this work, the multi-contrast MPI channel leakage is introduced and an analysis of the multi-contrast MPI system matrix properties is conducted to understand the source of the multi-contrast MPI channel leakage.
DDC Class
610: Medicine, Health
620: Engineering
Publication version
publishedVersion
Loading...
Name
IJMPI-Vol10-Iss1Suppl1-761.pdf
Type
Main Article
Size
342.45 KB
Format
Adobe PDF