  4. Deformation of press-fitted metallic resurfacing cups. Part 1: Experimental simulation
 
Options

Deformation of press-fitted metallic resurfacing cups. Part 1: Experimental simulation

Citation Link: https://doi.org/10.15480/882.1602
Publikationstyp
Journal Article
Date Issued
2006-02-01
Sprache
English
Author(s)
Jin, Zhongmin  
Meakins, Steven  
Morlock, Michael  
Parsons, Peter A.  
Hardaker, Catherine  
Flett, Magnus  
Isaac, Graham H.  
Institut
Biomechanik M-3  
TORE-DOI
10.15480/882.1602
TORE-URI
http://tubdok.tub.tuhh.de/handle/11420/1605
Journal
Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine  
Volume
220
Issue
2
Start Page
299
End Page
309
Citation
Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine 2 (220): 299-309 (2006)
Publisher DOI
10.1243/095441105X69150
Scopus ID
2-s2.0-33645232928
The interference press fit of a metallic one-piece acetabular cup employed for metal-on-metal hip resurfacing procedures was investigated experimentally under laboratory conditions in the present study, in particular regarding the cup deformation. Tests were carried out in cadavers as well as polyurethane foams of various grades with different elastic moduli to represent different cancellous bone qualities. The cadaver test was used to establish the most suitable configuration of the foam model representing realistic support and geometrical conditions at the pelvis. It was found that a spherical cavity, with two identical areas relieved on opposite sides, was capable of creating a two-point pinching action of the ischeal and ilial columns on the cup as the worst-case scenario. Furthermore, the cup deformation produced from such a two-point loading model with a grade 30 foam was similar to that measured from the cadaver test. Therefore, such a protocol was employed in subsequent experimental tests. For a given size of the outside diameter of the cup of 60 mm, the cup deflection was shown to be dependent largely on the cup wall thickness and the diametral interference between cup and prepared cavity at implantation. For a relatively thin cup with a wall thickness between 2.3 mm (equator) and 4 mm (pole) and with a modest nominal diametral interference of 1 mm, which corresponds to an actual interference of approximately 0.5 mm, the maximum diametral cup deflection (at the rim) was around 60 microm, compared with a diametral clearance of 80-120 microm between the femoral head and the acetabular cup, generally required for fluid-film lubrication and tribological performances. Stiffening of the cup, by both thickening and lateralizing by 1 mm, reduced the cup deformation to between 30 and 50 microm with actual diametral interferences between 0.5 and 1 mm.
Subjects
Arthroplasty, Replacement, Hip
Biocompatible Materials
Cadaver
Elasticity
Equipment Design
Equipment Failure Analysis
Hardness
Hardness Tests
Humans
In Vitro Techniques
Materials Testing
Metals
Pressure
Prosthesis Design
Surface Properties
Hip Prosthesis
Prosthesis Failure
DDC Class
610: Medizin
Lizenz
http://rightsstatements.org/vocab/InC/1.0/
Loading...
Thumbnail Image
Name

095441105x69150.pdf

Size

323.1 KB

Format

Adobe PDF

TUHH
Weiterführende Links
  • Contact
  • Send Feedback
  • Cookie settings
  • Privacy policy
  • Impress
DSpace Software

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science
Design by effective webwork GmbH

  • Deutsche NationalbibliothekDeutsche Nationalbibliothek
  • ORCiD Member OrganizationORCiD Member Organization
  • DataCiteDataCite
  • Re3DataRe3Data
  • OpenDOAROpenDOAR
  • OpenAireOpenAire
  • BASE Bielefeld Academic Search EngineBASE Bielefeld Academic Search Engine
Feedback