Respiration Extraction from Radar Heart Sound Measurements
Publikationstyp
Conference Paper
Date Issued
2019-07-01
Sprache
English
TORE-URI
Volume
2019
Start Page
6533
End Page
6536
Citation
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual Conference: 6533-6536 (2019-07-01)
Contribution to Conference
Publisher DOI
Scopus ID
Radar-based respiration measurement is susceptible to upper body movement in addition to the respiratory motion of the chest. This parasitic movement can only be canceled using a dual radar system from the front and back. However, the larger hardware effort could be avoided if a physiological parameter is measured that is influenced by respiration only but not by the movement. This could then be used to indirectly derive breathing. In this paper a method is presented, how the respiration can be deduced despite of upper body movements. To achieve this, a Six-Port interferometer is used to measure the heart sound envelogram of a test person from which subsequently the respiration can be reconstructed.