Options
Algorithmic channel decoupling for misaligned receive coils in magnetic particle imaging
Citation Link: https://doi.org/10.15480/882.4411
Publikationstyp
Journal Article
Date Issued
2022-03-21
Sprache
English
Author(s)
Gräser, Matthias
Institut
TORE-DOI
Volume
8
Issue
1, Suppl 1
Article Number
2203029
Citation
International Journal on Magnetic Particle Imaging 8 (1, Suppl 1): 2203029 (2022)
Publisher DOI
Scopus ID
Publisher
Infinite Science Publishing
Magnetic particle imaging (MPI) uses orthogonal receive coils aligned with the directions of the drive-field coils for signal detection. In case of misaligned receive coils, the particle magnetization couples differently into the receive coils such that a re-calibration involving the measurement of a new system function would be necessary. In this work, we propose a method for decoupling the channels algorithmically into the drive-field coil directions using an MPI transfer function matrix. In that way, system functions for different receive coil units can be reused.
DDC Class
004: Informatik
610: Medizin
More Funding Information
Research funding: The authors thankfully acknowledge the financial support by the German Research Foundation (DFG, grant number KN 1108/7-1 and GR 5287/2-1).
Publication version
publishedVersion
Loading...
Name
417-Article Text-2158-1-10-20220320.pdf
Size
656.86 KB
Format