A major issue for human-sizedMagnetic Particle Imaging scanners is the generation of sufficiently large magnetic gradient fields. By taking advantage of the field amplification properties of soft-iron, a considerable amount of power can be saved. In this work, an optimized selection and focus field generator is presented, that can generate flexible and high gradient fields at comparatively lowpower consumption. Coil spacing and possible field-free-point positions are similar to conventional MPI scanners designed with air coils, but with significantly less demands on infrastructure and cooling design. The optimization process is discussed and first field measurements are presented.
URI: http://hdl.handle.net/11420/13104
DOI: 10.15480/882.4545
ISSN: 2365-9033
Journal: International journal on magnetic particle imaging 
Institute: Biomedizinische Bildgebung E-5 
Document Type: Article
Project: Modellbasierte Parameteridentifikation in Magnetic Particle Imaging 
License: CC BY 4.0 (Attribution) CC BY 4.0 (Attribution)
Appears in Collections:Publications with fulltext

Files in This Item:
File Description SizeFormat
445-Article Text-2168-1-10-20220320.pdfVerlags-PDF1,09 MBAdobe PDFView/Open
Thumbnail
Show full item record

Page view(s)

42
checked on Jun 2, 2023

Download(s)

8
checked on Jun 2, 2023

Google ScholarTM

Check

Note about this record

Cite this record

Export

This item is licensed under a Creative Commons License Creative Commons