Abstract (english): Fretting corrosion as one reason for failure of modular hip prostheses has been associated with micromotion at the head taper junction. Historically the taper diameter was reduced to improve the range of motion of the hip joint. In combination with other developments, this was accompanied by increased observations of taper fretting, possibly due to the reduced flexural rigidity of smaller tapers. The aim of the study was to investigate how the flexural rigidity of tapers influences the amount of micromotion at the head taper junction. Three different stem and two different taper designs were manufactured. Experimental testing was performed using three different activity levels with peak loads representing walking, stair climbing and stumbling. The relative motion at the head-stem taper was measured in six degrees of freedom. Micromotion was obtained by subtraction of the elastic deformation derived from monoblock and finite element analysis. Less rigid tapers lead to increased micromotion between the head and stem, enlarging the risk of fretting corrosion. The influence of the stem design on micromotion is secondary to taper design. Manufacturers should consider stiffer taper designs to reduce micromotion within the head taper junction.
URI: http://hdl.handle.net/11420/2387
DOI: 10.15480/882.2216
ISSN: 1350-4533
Institute: Biomechanik M-3 
Type: (wissenschaftlicher) Artikel
Appears in Collections:Publications (tub.dok)

Files in This Item:
File Description SizeFormat
Haschke H 2019 MedEngPhys Influence of flexural rigidity AM.pdfAccepted Manuscript723,29 kBAdobe PDFEmbargoed until April 10, 2020    Request a copy
Show full item record

Page view(s)

40
Last Week
3
Last month
checked on May 25, 2019

Download(s)

12
checked on May 25, 2019

Google ScholarTM

Check

Export

Items in TORE are protected by copyright, with all rights reserved, unless otherwise indicated.