Publisher DOI: 10.1049/el.2018.6074
Title: Optimisation of inductive telemetry links in an implantable medical device for a wireless range of 25 cm
Language: English
Authors: John, Bibin 
Spink, Clemens 
Braunschweig, Markus 
Ranjan, Rajeev 
Schröder, Dietmar 
Koops, Andreas 
Woldt, Gregor 
Adam, Gerhard 
Krautschneider, Wolfgang 
Issue Date: 15-Nov-2018
Source: Electronics Letters 23 (54): 1315-1316 (2018-11-15)
Abstract (english): 
The optimisation of inductive telemetry link for increasing wireless range of a medical implant device is studied in detail. An active implant device using separate wireless power and data links at 2 MHz and 420 kHz has been fabricated and tested. A test setup consisting of RF power transmitter and data reader hardware has been implemented. Measurement results show that the fabricated device achieved a wireless range of 25 cm.