Publisher DOI: | 10.1049/el.2018.6074 | Title: | Optimisation of inductive telemetry links in an implantable medical device for a wireless range of 25 cm | Language: | English | Authors: | John, Bibin Spink, Clemens Braunschweig, Markus Ranjan, Rajeev Schröder, Dietmar Koops, Andreas Woldt, Gregor Adam, Gerhard Krautschneider, Wolfgang |
Issue Date: | 15-Nov-2018 | Source: | Electronics Letters 23 (54): 1315-1316 (2018-11-15) | Abstract (english): | The optimisation of inductive telemetry link for increasing wireless range of a medical implant device is studied in detail. An active implant device using separate wireless power and data links at 2 MHz and 420 kHz has been fabricated and tested. A test setup consisting of RF power transmitter and data reader hardware has been implemented. Measurement results show that the fabricated device achieved a wireless range of 25 cm. |
URI: | http://hdl.handle.net/11420/2695 | ISSN: | 0013-5194 | Journal: | Electronics letters | Institute: | Integrierte Schaltungen E-9 | Document Type: | Article | More Funding information: | BMBF, ESiMed [16M3201D] |