  4. In Vivo Quantification of the Apical Fixation Forces of Different Mitral Valved Stent Designs in the Beating Heart
 
Options

In Vivo Quantification of the Apical Fixation Forces of Different Mitral Valved Stent Designs in the Beating Heart

Publikationstyp
Journal Article
Date Issued
2015-05-01
Sprache
English
Author(s)
Pokorny, Saskia  
Huenges, Katharina  
Engel, A.  
Groß, Justus  
Frank, Derk  
Morlock, Michael  
Cremer, Jochen  
Lutter, Georg  
Institut
Biomechanik M-3  
TORE-URI
http://hdl.handle.net/11420/2970
Journal
Annals of biomedical engineering  
Volume
43
Issue
5
Start Page
1201
End Page
1209
Citation
Annals of Biomedical Engineering 5 (43): 1201-1209 (2015-05-01)
Publisher DOI
10.1007/s10439-014-1165-7
Scopus ID
2-s2.0-84939990384
The aim of this study was the quantification of the apical fixation forces of mitral valved stents for the first time. A test set-up was developed to measure the forces acting on the apical fixation of mitral valved stents. Twenty pigs received a self-expanding valved stent in the mitral position via a transapical approach in an off-pump procedure. Following stent deployment in the beating heart, the apical fixation forces were recorded following a standardized protocol and normalized. The apical fixation force of two different valved stent designs (AP: sole apical fixation, n = 10; SA: additional sub-annular fixation, n = 10) was compared to quantify the impact of the design change. Force measurement was successfully conducted in 18 of 20 animals (design AP: n = 10; design SA: n = 8). Evaluation showed higher mean forces and amplitude in the group with sole apical fixation (p = 0.046; Table 1). The impact of the additional sub-annular fixation was shown by a significant reduction of the force acting upon the apex of the heart. Quantification of these in vivo forces provides important information about the impact of different stent designs and might provide an objective parameter for stent performance evaluation in clinical application.
TUHH
Weiterführende Links
  • Contact
  • Send Feedback
  • Cookie settings
  • Privacy policy
  • Impress
DSpace Software

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science
Design by effective webwork GmbH

  • Deutsche NationalbibliothekDeutsche Nationalbibliothek
  • ORCiD Member OrganizationORCiD Member Organization
  • DataCiteDataCite
  • Re3DataRe3Data
  • OpenDOAROpenDOAR
  • OpenAireOpenAire
  • BASE Bielefeld Academic Search EngineBASE Bielefeld Academic Search Engine
Feedback