Title: Deformation of acetabular press-fit cups: influence of design and surgical factors
Language: English
Authors: Messer-Hannemann, Philipp 
Campbell, Graeme Michael 
Morlock, Michael 
Keywords: hip arthroplasty;press-fit;cup deformation;primary stability;contact condition
Issue Date: 13-Jul-2019
Publisher: Elsevier Science
Source: Clinical Biomechanics (69): 96-103 (2019-10)
Journal or Series Name: Clinical biomechanics 
Abstract (english): Background: Deformation of acetabular cups when press-fitted into an undersized cavity is inevitable due to the inhomogeneous stiffness of acetabular bone. Thinner cups or screw holes might increase the risk of high cup deformation. The aim of this study was to examine the influence of cup design and liner assembly on the deformation response during cup implantation. Methods: Acetabular cups with different designs were implanted into polyurethane foam models simulating the anatomical situation with nominal press-fits of 1mm and without nominal press-fits (line-to-line). Deformations were determined using a tactile coordinate measuring machine. A 3D laser scanner was used to determine the contact conditions at the cup-cavity interface. Polyethylene and ceramic liners were assembled to the implanted cups and the influence of the insertion on the deformation response evaluated. Fixation strength of the cups was determined by push-out testing. Findings: Cup deformation increased with smaller wall thickness (P < 0.037) and screw holes (P < 0.001). Insertion of ceramic liners reduced the deformation (P < 0.001), whereas polyethylene liners adapted to the deformation of the implanted cups (P > 0.999). Thin-walled cups exhibited a higher fixation strength for similar implantation forces (P = 0.011). Interpretation: Thin-walled cups achieved higher fixation strengths and might be more bone-preserving. However, in combination with screw holes and high press-fit levels, wall thickness should be considered carefully to avoid excessive cup deformations leading to potential complications during liner assembly. Line-to-line insertion of thin-walled cups should be accompanied with a rough surface coating to minimize the loss of fixation strength due to the low press-fit fixation.
URI: http://hdl.handle.net/11420/3058
DOI: 10.15480/882.2363
ISSN: 1879-1271
Institute: Biomechanik M-3 
Type: (wissenschaftlicher) Artikel
Appears in Collections:Publications (tub.dok)

Files in This Item:
File Description SizeFormat
Messer-Hannemann P 2019 ClinBio Deformation of acetabular press-fit cups - Influence of design and surgical factors.pdfaccepted manuscript1,66 MBAdobe PDFEmbargoed until October 1, 2020    Request a copy
Show full item record

Page view(s)

113
Last Week
22
Last month
checked on Aug 25, 2019

Download(s)

11
checked on Aug 25, 2019

Google ScholarTM

Check

Export

This item is licensed under a Creative Commons License Creative Commons