Publisher DOI: 10.1016/j.jbiomech.2019.109424
Title: Micromotion at the head-stem taper junction of total hip prostheses is influenced by prosthesis design-, patient- and surgeon-related factors
Language: English
Authors: Falkenberg, Adrian 
Biller, Sara 
Morlock, Michael 
Huber, Gerd 
Keywords: Assembly force; Head length; Head-stem taper junction; Micromotion; Taper surface topography
Issue Date: 2-Jan-2020
Source: Journal of Biomechanics (98): 109424 (2020-01-02)
Abstract (english): 
Taper junctions of modular hip prostheses are susceptible to fretting and crevice corrosion. Prevalence and significance increase for cobalt-chromium heads assembled on titanium-alloy stems. Retrieval and in-vitro studies have identified micromotion between the taper components to accelerate the corrosion process. The aim of this study was to identify the most critical factors contributing to increased micromotion, which is most likely influenced by design-, patient- and surgeon-related aspects. Micromotion between head and stem taper surfaces was measured for different taper surface topographies and load orientations. Consecutive visual images were recorded through windows in the head component. By image matching analysis the local micromotions at the taper junction between head and stem tapers were determined. To extend the findings to taper regions not visible through the windows, finite element models were generated. The models were further utilized to investigate the influence of head length, taper angle difference and assembly force on micromotion. Significantly higher micromotion (+20%) was found under varus loading (7.1 µm) in comparison to valgus loading (5.9 µm). Smooth and microgrooved stem tapers exhibited equal amounts of micromotion. The numerical model revealed head tilting and recurring taper contact changes in terms of cyclic engagement/disengagement during the loading sequences. Especially long heads (+240%) and low assembly forces (+53%) were found to substantially increase micromotion (from 2.7 µm to 9.3 µm and from 4.1 µm to 8.8 µm, respectively). This study accentuates the susceptibility of taper junctions to a variety of factors, which need to be appreciated in preoperative planning and surgical procedure to reduce the amount of micromotion and such minimize the risk of critical corrosion.
URI: http://hdl.handle.net/11420/4244
ISSN: 0021-9290
Journal: Journal of biomechanics 
Institute: Biomechanik M-3 
Document Type: Article
Appears in Collections:Publications without fulltext

Show full item record

Page view(s)

106
Last Week
1
Last month
0
checked on Jun 8, 2023

SCOPUSTM   
Citations

16
Last Week
0
Last month
1
checked on Jun 30, 2022

Google ScholarTM

Check

Add Files to Item

Note about this record

Cite this record

Export

Items in TORE are protected by copyright, with all rights reserved, unless otherwise indicated.