  4. Quantitative Analysis of 3D Artery Volume Reconstructions Using Biplane Angiography and Intravascular OCT Imaging
 
Options

Quantitative Analysis of 3D Artery Volume Reconstructions Using Biplane Angiography and Intravascular OCT Imaging

Publikationstyp
Conference Paper
Date Issued
2019-07
Sprache
English
Author(s)
Latus, Sarah  orcid-logo
Neidhardt, Maximilian  
Lutz, Matthias  
Gessert, Nils Thorben  
Frey, Norbert  
Schlaefer, Alexander  
Institut
Medizintechnische Systeme E-1  
TORE-URI
http://hdl.handle.net/11420/4307
Start Page
6004
End Page
6007
Citation
Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC): 6004-6007 (2019-07)
Contribution to Conference
Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC 2019)  
Publisher DOI
10.1109/EMBC.2019.8857712
Scopus ID
2-s2.0-85077872679
Diameter and volume are frequently used parameters for cardiovascular diagnosis, e.g., to identify a stenosis of the coronary arteries. Intra-vascular OCT imaging has a high spatial resolution and promises accurate estimates of the vessel diameter. However, the actual images are reconstructed from A-scans relative to the catheter tip and imaging is subject to rotational artifacts. We study the impact of different volume reconstruction approaches on the accuracy of the vessel shape estimate. Using X-ray angiography we obtain the 3D vessel centerline and the 3D catheter trajectory, and we propose to align the A-scans using both. For comparison we consider reconstruction along a straight line and along the centerline. All methods are evaluated based on an experimental setup using a clinical angiography system and a vessel phantom with known shape. Our results illustrate potential pitfalls in the estimation of the vessel shape, particularly when the vessel is curved. We demonstrate that the conventional reconstruction approaches may result in an overestimate of the cross-section and that the proposed approach results in a good shape agreement in general and for curver segments, with DICE coefficients of approximately 0.96 and 0.98, respectively.
TUHH
Weiterführende Links
  • Contact
  • Send Feedback
  • Cookie settings
  • Privacy policy
  • Impress
DSpace Software

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science
Design by effective webwork GmbH

  • Deutsche NationalbibliothekDeutsche Nationalbibliothek
  • ORCiD Member OrganizationORCiD Member Organization
  • DataCiteDataCite
  • Re3DataRe3Data
  • OpenDOAROpenDOAR
  • OpenAireOpenAire
  • BASE Bielefeld Academic Search EngineBASE Bielefeld Academic Search Engine
Feedback