  4. The effect of body weight on the choice of material for the bearing couple in artificial hip joints
 
Options

The effect of body weight on the choice of material for the bearing couple in artificial hip joints

Publikationstyp
Journal Article
Date Issued
2014-01-15
Sprache
English
Author(s)
Hothan, Arne  
Morlock, Michael  
Hoenig, Elisa  
Institut
Biomechanik M-3  
TORE-URI
http://hdl.handle.net/11420/5888
Journal
Seminars in arthroplasty  
Volume
24
Issue
4
Start Page
218
End Page
239
Citation
Seminars in Arthroplasty 4 (24): 218-239 (2013)
Publisher DOI
10.1053/j.sart.2014.01.008
Scopus ID
2-s2.0-84896827639
Publisher
Elsevier
The effect of body weight on the success of arthroplasty replacement of the hip joint is a controversial issue, particularly the question of whether the choice of bearing couple used should be adapted to individual patient characteristics. As part of this study, we carried out a literature-based meta-analysis of clinical results that referred to patient weight. The friction of various material combinations under different joint forces was also experimentally determined. Clinically, there are reports of unfavorable implant positioning, increased rate of dislocations, and increased noise developments for obese patients. The results of the friction measurements do not suggest a material choice adapted to body weight. For obese patients, therefore, the same argument applies as for patients with normal weight; provided a correct implantation situation can be ensured, ceramic-on-ceramic bearing couples should be favored from a tribological point of view. If the correct implant position - for example, based on the more complicated implant conditions in obese patients - cannot be achieved, a hard-soft bearing couple with a ceramic head appears to be preferable to a hard-hard bearing couple.
Subjects
Arthroplasty
Body weight
Ceramic
Friction
Material
Metal
Polyethylene
DDC Class
570: Biowissenschaften, Biologie
610: Medizin
TUHH
Weiterführende Links
  • Contact
  • Send Feedback
  • Cookie settings
  • Privacy policy
  • Impress
DSpace Software

Built with DSpace-CRIS software - Extension maintained and optimized by 4Science
Design by effective webwork GmbH

  • Deutsche NationalbibliothekDeutsche Nationalbibliothek
  • ORCiD Member OrganizationORCiD Member Organization
  • DataCiteDataCite
  • Re3DataRe3Data
  • OpenDOAROpenDOAR
  • OpenAireOpenAire
  • BASE Bielefeld Academic Search EngineBASE Bielefeld Academic Search Engine
Feedback