    Corrosion of the Head-Stem Taper Junction—Are We on the Verge of an Epidemic?: Review Article
    Background: The modular head taper junction has contributed to the success of total hip arthroplasty (THA) greatly. Taper corrosion and wear problems reported for large and extra-large metal-on-metal bearings as well as for bi-modular THA stems have cast doubt on the benefit of the taper interface. Presently, corrosion problems are being reported for nearly all kinds of artificial hip joints incorporating metal heads, questioning taper connections in general. Questions/purposes: This study aimed to review the mechanical and electrochemical relationships that may lead to taper corrosion, which have been reported more commonly in recent literature, and to also review the contribution of patient characteristics and surgical techniques involved in taper assembly that may contribute to the problem. Methods: The search criteria “(corrosion) AND (hip arthroplasty) AND (taper OR trunnion)” and “(hip arthroplasty) AND ((pseudotumor) OR (pseudo-tumor))” in PubMed and the JAAOS were used for the literature search. In addition, the arthroplasty registers were considered. Results: Most studies acknowledge the multifactorial nature of the problem but concentrate their analysis on taper and implant design aspects, since this is the only factor that can be easily quantified. The sometimes conflicting results in the literature could be due to the fact that the other two decisive factors are not sufficiently considered: the loading situation in the patient and the assembly situation by the surgeon. All three factors together determine the fate of a taper junction in THA. There is no single reason as a main cause for taper corrosion. The combined “outcome” of these three factors has to be in a “safe range” to achieve a successful long-term taper fixation. Conclusion: No, this is not the beginning of an epidemic. It is rather the consequence of disregarding known mechanical and electrochemical relationships, which in combination have recently caused a more frequent occurrence—and mainly reporting—of corrosion issues.
    Publicationtype: Review Article
      70
  • Some of the metrics are blocked by your 
    Item type:Publication,
    Quantification of material loss from the neck piece taper junctions of a bimodular primary hip prosthesis. A retrieval study from 27 failed Rejuvenate bimodular hip arthroplasties
    The early failure and revision of bimodular primary total hip arthroplasty prostheses requires the identification of the risk factors for material loss and wear at the taper junctions through taper wear analysis. Deviations in taper geometries between revised and pristine modular neck tapers were determined using high resolution tactile measurements. A new algorithm was developed and validated to allow the quantitative analysis of material loss, complementing the standard visual inspection currently used. The algorithm was applied to a sample of 27 retrievals (in situ from 2.9 to 38.1 months) of the withdrawn Rejuvenate modular prosthesis. The mean wear volumes on the flat distal neck piece taper was 3.35 mm(3) (0.55 to 7.57), mainly occurring in a characteristic pattern in areas with high mechanical loading. Wear volume tended to increase with time to revision (r² = 0.423, p = 0.001). Implant and patient specific data (offset, stem size, patient's mass, age and body mass index) did not correlate with the amount of material loss observed (p > 0.078). Bilaterally revised implants showed higher amounts of combined total material loss and similar wear patterns on both sides. The consistent wear pattern found in this study has not been reported previously, suggesting that the device design and materials are associated with the failure of this prosthesis.
    Publicationtype: Journal Article
      130